Biochemical System
SANTHIA 240 Premium
Innovative device for liquid chemistry.
- Cuvettes: 80 washable BIONEX cuvettes
- Reaction volume: 190-350 uL
- Reaction time: maximum incubation + reading time 800 seconds
- Reaction temperature: 37 C, heating module adjustable from room temperature up to 42 C
- Optical measurement: 8 fixed wavelengths, 340-700 nm
- Optical source: 1 halogen lamp, 6V 10W, with extended UV emission
- Water consumption: up to 1-2 L/hour, depending on workload
- Cuvette washing: 8-step washing process for each cuvette
- System type: Random Access/STAT
- Speed: up to 220 tests/hour, fixed cycle
- Analytical methods: endpoint, bichromatic endpoint, differential endpoint, differential endpoint sample blank, fixed time, kinetic, kinetic bichromatic
- Calibration: up to 8 points per test method, including linear, spline, and logit-log
- Sample carrier: removable holders, 32 positions
- Sample vessel: tubes 12-13 mm x 75 mm or cups with adapter, 1 mL
- Sample volume: 2-300 uL
- Dilution: dilution in the needle when the sample volume of the method allows; automatic pre-dilution in the reaction cuvette up to 1:100
- Sample barcode: integrated barcode reader
- Bottle volume: 40 mL, 25 mL, 15 mL, 2 mL
- Reagent volume: 5-350 uL
- Reagent storage: on-board reagent cooling
- Reagent carrier: removable holders, 32+32 bottles
- Reagent barcode: integrated barcode reader
- Dead volume: no dead volume
- Communication: USB port
- Result printing: individual test, final sample, worksheet, method and QC
- Power supply: 100-240 V, 50/60 Hz
- Dimensions: 54 x 60 x 64 cm (W x D x H)
- Weight: 40 kg